Terminated

AmniSure ROM Test, Fetal Membrane Rupture Test for In-Vitro Diagnostic Use, Model: FMRT-1, AmniSure International LLC, Cambridge, MA 02138

FDA Recall Z-0120-2008 · Posted 2008-01-19

Recall StatusTerminated
Product CodeJJX
Date Initiated2007-10-03
Date Terminated2008-02-06
Product Quantity319,976 units

Reason for Recall

Weak true postive/false negative results: weak true positive or false negatives in ruptured fetal membranes may occur due to a "hook effect"

Firm Action

AmniSure notified users by Medical Device Correction (Recall) letter dated October 1, 2007 issued via US Mail. The letter provides additional labeling instructions and a checklist for trouble shooting, including a diluation procedure that allows users to determine and solve the problem of the hook effect with this device.

Root Cause

Device Design

Distribution Pattern

Worldwide, USA, Australia, Austria, Italy, Korea, Bulgaria, Russia, Spain, Kuwait, Switzerland, Luexemburg, Ireland, Denmark, Netherlands, Jordan, Israel, Chile, UK, Singapore, Basil, New Zealand, Turkey, France, Greece, and Equador