Terminated

Disposable Smooth Interior Tubing Material: Polyethylene EVA copolymer. This tubing is supplied non-sterile P/N: 1024920 Rev. 01 Low temperature smooth interior tubing is a disposable pulmonary equipment interconnect used in ventilation and aesthesia machines.

FDA Recall Z-0120-2019 · Posted None

Recalling FirmA M Systems Inc
Recall StatusTerminated
Product CodeBYX
Date Initiated2018-08-10
Date Terminated2021-05-13
Product Quantity610

Reason for Recall

Some of the Disposable Smooth Interior Tubing may pull apart when stretched.

Firm Action

On 08/10/2018, the firm, A-M Systems, sent an "URGENT: Notification of Product Recall/Correction" letter to customers via email informing them that it was determined that some of the Disposable Smooth Interior Tubing may pull apart when stretched. Customers were advised to provide the firm with the number of items distributed from the affected lots and the number of items still remaining in their inventory from these lots so that the firm could accurately track the status of this correction action. The customers are instructed to please use the enclosed form and self-addressed envelope to return the information to the firm immediately. Once the customers have provided the firm with an accurate count of items distributed and items remaining in the customer's control please destroy or discard any remaining product in your possession from these lot numbers. If you have any questions, please contact Quality Manager at Tel: 360.683.8300 x1302 or Email: [email protected].

Root Cause

Process control

Distribution Pattern

US Distribution to states of: OH, PA, and NC.

Official Source

View FDA Recall →