MEDLINE ReNewal Harmonic ACE+7 Shears w/ Adv. Hemostasis Use Only w/ GEN11 SW v2018-1 and Lower, REF HARH45R. surgical instrument.
FDA Recall Z-0120-2024 · Posted 2023-10-16
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeNLQ
Date Initiated2023-08-03
Product Quantity18 units
Reason for Recall
Various reprocessed Covidien LigaSure and Ethicon Harmonic devices may have an incomplete seal on the packaging tray.
Firm Action
Medline issued a recall notice to its consignees by email on 08/03/2023. The notice explained the problem and the potential risk. Those consignees who have transferred possession of the affected product were directed to notify their customers. The firm is seeking the return of the affected product.
An error in the model number was noted (missing an "R" at the end) and the firm issued an updated letter on 09/19/2023 with the corrected number.
Root Cause
Packaging
Distribution Pattern
US
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.