Monode, Shortwave Accessory Model #: 02200002 Used in conjunction with the Intelect Shortwave Diathermy 100 and Intelect Shortwave Diathermy 400 units. For the clinical application of electromagnetic energy in the radio frequency of 27.12 megahertz to generate deep heat within body tissues.

FDA Recall Z-0121-2016 · Posted 2015-10-15

Recalling FirmDJO, LLC
Recall StatusTerminated
Product CodeIMJ
Date Initiated2015-08-21
Date Terminated2016-10-24
Product Quantity2887

Reason for Recall

Overheating and burning of the wire insulation inside the applicator.

Firm Action

The firm sent out two customer notification letters; one to distributors and one to direct users. Letters were sent September 1, 2015 via USPS certified return receipt. Distributors were asked to obtain a list of their customers with contact information that have purchase or received the monode accessory or Intelect Shortwave device that included the monode as an accessory item. Once the list has been assembled the firm requests that it be forwarded to DJO Global Chattanooga Product Support by phone 877-733-4888, or by e-mail at [email protected]. The firm's notification states that customers will be sent an upgraded monode accessory along with a pre-paid shipper for the return of the current used unit. Any questions the letter states to contact Product Support per the contact information above or William Fisher, Global Regulatory Compliance at 760-734-3126. Letters sent to direct users received the same instruction of returning the product for a replacement. Any questions the letter states to call 877-733-4888 or email at [email protected].

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of WY, OH, WA, NC, FL, NE, GA, KS CA, MO, IN, MA, IL, OK, AZ, TX, AR, UT, NJ, WI, PA, SC, WV, NY, MI, MN, VA, AK, TN, OR, NC, NH, KY, CO, NV, ID and the countries of El Salvador, China, Ecuador, South Korea, Chile, Israel, Peru, Nicaragua, Bolivia, Indonesia