ESPOCAN Combined Spinal and Epidural Anesthesia Tray, ESPOCAN 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN 25 Ga. x 5 in. (12.7 cm) Spinal Needle with Marking and Centering Sleeve, and 19 Ga. Closed Tip PERIFIX FX Springwound Epidural Catheter. Product Code: ES1725KFX. Product Description: ES1725KFX ESPOCAN SPINAL/EPID. TRAY.

FDA Recall Z-0122-2026 · Posted 2025-10-09

Recalling FirmB BRAUN MEDICAL INC
Recall StatusOpen, Classified
Product CodeOGE
Date Initiated2025-09-18
Product Quantity10,130 units

Reason for Recall

Potential for the lid of the catheter connector to be in the incorrect position.

Firm Action

B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distributed, notify customers, post the notification where affected units are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form." Additionally, if consignees have identified affected units in inventory, they may order one case (50 pieces) of affected sterilized single use catheter connectors to have on hand.

Root Cause

Process control

Distribution Pattern

Nationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.