Hydrocision SpineJet XL MIS 30 degree Handpieces, for cutting, ablation, and removal of soft tissue, hard tissue, and bone. Catalog Number 58663
FDA Recall Z-0124-2007 · Posted 2006-11-02
Advertisement
Recalling FirmHydrocision, Inc.
Recall StatusTerminated
Product CodeHRX
Date Initiated2006-09-07
Date Terminated2006-12-06
Product Quantity40 total
Reason for Recall
Distal tip may become detached from the device
Firm Action
Hydrocision contacted customers and sales reps by email on 9/6/06 advising to users to discontinue use and return recalled units.
Root Cause
Process change control
Distribution Pattern
Nationwide.