Sofia SARS Antigen FIA Package Insert The Sofia SARS Antigen FIA is a lateral flow immunofluorescent sandwich assay that is used with the Sofia and Sofia 2 instrument intended for the qualitative detection of the nucleocapsid protein antigen from SARS-CoV-2 in nasopharyngeal (NP) and nasal (NS) swab specimens directly from individuals who are suspected of COVID-19 by their healthcare provider within the first five days of the onset of symptoms.

FDA Recall Z-0124-2021 · Posted None

Recalling FirmQuidel Corporation
Recall StatusTerminated
Product CodeQKP
Date Initiated2020-09-17
Date Terminated2022-08-09
Product Quantity10537

Reason for Recall

Instructions for use in the package insert for SARS Antigen FIA test were updated to remove all references to a specific viral transport medium product as well as the use of other viral transport media. Use of the specific viral transport medium products may result in False-positive results. Directly testing from specimens (NS or NP swab) is recommended.

Firm Action

On 09/17/20, "FIELD CORRECTIVE ACTION CORRECTION" notices dated 09/16/20 were emailed to customers. The firm called customers over the phone to convey the correction information to some customers. Customers were asked to take the following actions: 1) Discontinue the use of affected package inserts. 2) Access the updated instructions for use labeling which can be located by going to quidel.com/docs and selecting the Sofia SARS Antigen FIA Package Insert document. 3) In the event that your facility has used Remel viral transport media products (e.g., M4, M4RT, M6), please contact Quidel Technical Support to discuss your results. 4) Confirm receipt of this notice by returning a signed copy to [email protected] Please feel free to contact Technical Support at [email protected] or by calling 800.874.1517 (in the U.S.) with any questions or concerns that you may have regarding this issue.

Root Cause

Device Design

Distribution Pattern

U.S. Nationwide Distribution: NY, KS, TX, AL, FL, MN, CA, IL, MA, MS, PA, GA, NH, MO, VA, MI, OR, WV, OK, NV, CT, TN, MD, LA, AR, HI, NJ, KY, VT, IA, RI, NM, OH, IN, AK, SC, NC, AZ, SD, NE, WI, and.CO.