Medtronic Sofamor Danek COLORADO 2" SPINAL SYSTEM Break-Off Nut, non-sterile, RX; Catalog Number 8634111, Medtronic, Memphis, TN 38138.
FDA Recall Z-0125-2008 · Posted 2008-02-07
Advertisement
Recalling FirmWarsaw Orthopedic, Inc. dba Medtronic SOFAMOR DANEK
Recall StatusTerminated
Product CodeKWP
Date Initiated2007-09-10
Date Terminated2009-07-14
Product Quantity2,569
Reason for Recall
Assembly may be compromised: Two thread grooves are missing on the nut, which may cause damage to the bone screw and result in the need for the bone screw to be replaced.
Firm Action
Medtronic Sofamor Danek notified consignees by Urgent Device Recall letter dated 9/10/07 to immediately cease using the product and to return the devices with a Product Recall Questionnaire, recording quantity of devices and any patient injuries.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution - USA, Australia, Hong Kong and the Netherlands.