Terminated

Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device delivered via catheter to treat descending abdominal aortic aneurysms (AAAs).

FDA Recall Z-0125-2015 · Posted 2014-11-03

Recalling FirmTrivascular, Inc
Recall StatusTerminated
Product CodeMIH
Date Initiated2014-09-20
Date Terminated2014-12-04
Product Quantity0 Devices

Reason for Recall

TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.

Firm Action

Urgent Field Safety Notice letter was sent to all consignees on 9/19/14 via trackable mail system. Users are notified that TriVascular has identified select lots of Ovation Prime 29mm Aortic Body stent grafts that may be associated with rapid emptying of the fill polymer syringe when it is connected to the Autoinjector during the fill polymer injection step of the primary implant procedure.

Root Cause

Under Investigation by firm

Distribution Pattern

International Distribution Only -- Canada, Turkey, Austria, Ireland, Belgium, Italy, Switzerland, Spain, France, and Cyprus.

Official Source

View FDA Recall →