AMPLATZER Vascular Plug, Manufactured by AGA Medical Corporation , 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA. Sterile EO. Single Use Only. Rx Only. Do not use open or damaged packages. Store in a cool, dry place. The AMPLATZER Vascular Plug is a self-expandable, cylindrical device made from a Nitinol wire mesh. The device is secured on both ends with platinum marker bands. A stainless steel micro screw is welded to one of the platinum marker bands, which allows attachment to the 135 cm long deliver cable. The AMPLATZER Vascular Plugs are made in various sizes ranging from 4 to 16 mm in 2 millimeter increments.

FDA Recall Z-0127-06 · Posted 2005-11-05

Recall StatusTerminated
Product CodeKRD
Date Initiated2005-09-01
Date Terminated2006-11-13
Product Quantity81

Reason for Recall

Five lots of AMPLATZER Vascular Plugs were sterilized in a load that was not validated for the AMPLATZER Vascular Plug. However, only three of those five lots were distributed.

Firm Action

Urgent Recall letters, dated 09/01/05, were sent to affected customers via email, facsimile and with follow-up delivery via USPS. OUS customers were notified via Federal Express. The letter described the situation, the lots being recalled and included a response form that was to be faxed back to AGA Medical.

Root Cause

Other

Distribution Pattern

OUS countries include: Australia, Germany, Italy, Korea, Malasia, Slovakia, Spain, Switzerland, Thailand, United Kingdom