Dimension Vista HbA1c

FDA Recall Z-0127-2015 · Posted 2014-10-24

Recall StatusTerminated
Product CodeLCP
Date Initiated2014-08-14
Date Terminated2015-08-10
Product Quantity20043

Reason for Recall

In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t

Firm Action

An Urgent Medical Device Correction, dated August 2014, was provided to all Dimension Vista HbA1c customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution: US (nationwide) and country of: Canada.