4FR Single Lumen Pressure Injectable PICC; intravascular therapeutic catheter, for long-term use greater than 30 days.
FDA Recall Z-0127-2016 · Posted 2015-10-15
Advertisement
Recalling FirmArrow International Inc
Recall StatusTerminated
Product CodeLJS
Date Initiated2015-09-18
Date Terminated2020-06-25
Product Quantity31590
Reason for Recall
Labeling inaccuracy in the Instructions for Use (IFU) and Patient Card. The IFU and Patient Card with these products state the PICC can withstand up to ten (10) pressure injections, however, the PICC has only been tested to withstand five (5) pressure injections.
Firm Action
An urgent medical device notification, dated September 22, 2015, was sent to direct accounts to alert them about the issue, the potential risks, and provide instructions for returning the completed response form and affected devices.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide distribution. US Nationwide, Australia, Canada, and New Zealand.