Terminated

Bari Maxx II, KCI''s BariMaxx II bariatric bed, P/N 310611, November 2001, KCI The Clinical Advantage, P. O. Box 659508, San Antonio, Texas 78265-9508, Call Toll Free 1-800-531-5346, FAX (210) 308-3998, Http://www.kci1.com

FDA Recall Z-0129-06 · Posted 2005-11-08

Recalling FirmKinetic Concepts, Inc
Recall StatusTerminated
Product CodeFNL
Date Initiated2004-08-16
Date Terminated2005-11-08
Product Quantity93 beds

Reason for Recall

Design change to uppgrade reliability of powered adjustable bed.

Firm Action

All consignees were notified with a technical bulletin explaining the inspection and addition of washers to the set-screws.

Root Cause

Other

Distribution Pattern

Nationwide

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑