American Medical Systems, Inc., (AMS) GreenLight HPS Fiber Optic (ADDStat) Part Number 0010-2090B. Fiber optic laser for treatment of BPH (Benign Prostatic Hyperplasia). AMS Innovative Center, San Jose, CA 95134
FDA Recall Z-0131-2008 · Posted 2008-02-26
Advertisement
Recalling FirmAMS Innovative Center - San Jose
Recall StatusTerminated
Product CodeGEX
Date Initiated2007-06-18
Date Terminated2010-03-23
Product Quantity611 units
Reason for Recall
Control knob break: The knob used to finely control the direction of the laser energy may break.
Firm Action
American Medical Systems, Inc. (AMS) issued Important: Medical Device Recall AMS Greenlight ADDStat Fibers notification letters to customers who have purchased the product on June 28, 2007. The customer letter requests recipients that the product use be discontinued, and the fibers be returned. The firm will send replacements.
Root Cause
Process design
Distribution Pattern
Worldwide Distribution - USA and Singapore.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.