18 gage XTW needles, sold as a component in SafeSheath Hemostatic Tear-away Introducer System with Infusion Side Port kits or sold Bulk Non Sterile (BNS).
FDA Recall Z-0131-2009 · Posted 2008-11-03
Advertisement
Recalling FirmThomas Medical Products Inc
Recall StatusTerminated
Product CodeDYB
Date Initiated2008-09-12
Date Terminated2009-05-06
Product Quantity246,585 needles
Reason for Recall
Difficulty inserting guidewire through introducer needle. Needle hub is not tapered down to the needle cannula.
Firm Action
On 9/11/08, Thomas Medical Products issued letters to the distributors informing them of the problem, to cease distribution and await further instructions. Additional instructions to the distributors were to notify all their account via an advisory notice and request that the user not use the needle supplied with the kit and to replace with another commercially available sterile 18 gage introducer needle.
Root Cause
Process control
Distribution Pattern
Nationwide distribution via distributors in CA, MA, and MN.