Acrobat V Vacuum Off-Pump System OM-9100S Acrobat SUV Vacuum Off-Pump System OM-9000S Axius Xpose 3 Access Device XP-3000 XPOSE 4 Access Device XP-4000 Acrobat V Vacuum Off-Pump System Each Acrobat V Vacuum Off-Pump system is intended for use during performance of cardiac surgical procedures through a sternotomy incision approach.

FDA Recall Z-0132-2015 · Posted 2014-10-24

Recall StatusTerminated
Product CodeMWS
Date Initiated2014-08-13
Date Terminated2016-03-30
Product Quantity16,532 units (19,270 units international)

Reason for Recall

An internal finding identified the presence of pinholes/cuts in some of the Tyvek pouches of the vacuum tubing sets included as an accessory to Acrobat V Vacuum Off-Pump System, Acrobat SUV Vacuum Off-Pump System, Axius XPOSE 3 Access Device and XPOSE 4 Access Device.

Firm Action

Maquet Inc. issued Recall letters/Acknowledgement forms dated via Fed-Ex on 8/13/2014.

Root Cause

Package design/selection

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: ARGENTINA, AUSTRIA, BANGLADESH, BELARUS, BRAZIL, BRUNEI, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ECUADOR, GERMANY, FINLAND, FRANCE, GREECE, HONG KONG, INDICA, INDONESIA, ISRAEL, ITALY, JAPAN, MEXICO, MALAYSIA, NETHERLANDS, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAK REPUBLIC, SOUTH KOREA, SPAIN, SWITZERLAND, TAIWAN, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM AND VIETNAM.