The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

FDA Recall Z-0132-2026 · Posted 2025-10-10

Recalling FirmInsulet Corporation
Recall StatusOpen, Classified
Product CodeQRX
Date Initiated2025-09-11
Product Quantity174,013 units

Reason for Recall

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Firm Action

Insulet notified both healthcare professionals and users on about 09/11/2025 via URGENT: Medical Device Correction letter. Instructions included that an update to the application will be pushed to all devices and should be implemented by users. Users were also requested to confirm receipt and acknowledgement of the notification but selecting the "acknowledgement button" in the email.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution.