Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product Code # 2M8064. The Flo-Gard Volumetric Infusion Pumps are an electromechanical pump used for intravenous infusion of fluids at user-selected rates.

FDA Recall Z-0134-2012 · Posted 2011-10-28

Recall StatusTerminated
Product CodeFRN
Date Initiated2011-08-26
Date Terminated2011-10-31
Product Quantity1 unit

Reason for Recall

The Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test.

Firm Action

A Baxter Healthcare technician initially notified the sole customer on August 26, 2011 of the defective product and requested that the product be quarantined. On September 14, 2011 a technician visited the site to correct the problem observed on the Flo-Gard Infusion Pump. All required tests were performed and the results were found acceptable.

Root Cause

Employee error

Distribution Pattern

Distributed only in Puerto Rico.