Terminated

THSCF-35-260-3-AUS2, (Cook Vascular Wire Guide, PTFE coated curved Amplatz ultra stiff catheter exchange, Diameter .035 Inch, length 260 cm, curve radius 3 mm, Cook Incorporated, Bloomington, IN 47402 - not on label)

FDA Recall Z-0135-2008 · Posted 2008-02-06

Recalling FirmCook, Inc.
Recall StatusTerminated
Product CodeDQX
Date Initiated2007-08-15
Date Terminated2008-04-21
Product Quantity33

Reason for Recall

The device is normally sold as sterile, however this lot was not sterilized and the device is not adequately labeled, as the label does not state whether the device is sterile or non-sterile and lacks the manufacturer's name and address.

Firm Action

Cook Incorporated notified consignees by Urgent Product Recall letter dated 8/15/07 instructing them to return the product.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - USA, Australia, India, Trinidad and Tobago.