Terminated

Monarch Inflation Device and Fluid Dispensing Syringe - Latex Free MERIT MEDICAL. Sterile if package is unopened or undamaged. Part # IN2125/B .

FDA Recall Z-0137-04 · Posted 2003-11-19

Recall StatusTerminated
Product CodeDXT
Date Initiated2003-10-01
Date Terminated2004-02-02
Product Quantity1997 units

Reason for Recall

There is a potential for the PETG blister trays to bond together with varying degrees of attachment. The potential for non-sterility exists when the attached trays are seperated by force and holes or cracks are generated in the seperated trays.

Firm Action

Merit notified sales representatives, domestic and international accounts and dealers immediately by letter dated October 3, 2003. The manufacturer requests that the use of the affected inventory should immediately cease and that on-hand units should be promptly isolated from general inventory for simultaneous inspection by a site representative and a Merit sales representative.

Root Cause

Other

Distribution Pattern

Distributed nationwide. International distribution includes Canada, Japan, Korea, England and France.