Terminated

Vitros ECi Immunodiagnostic System, sold as Cat. #863 3893; Cat. #192 2814 (Vitros ECiQ), and Cat. #680 1059 (recertified Vitros ECi) when using Vitros Anti-HBc IgM Reagent Packs and Anti-HAV IgM Reagent Packs. NOTE: These 3 catalog numbers of Vitros ECi represent the original unit, an updated unit with cosmetic changes only, and refurbished units.

FDA Recall Z-0137-06 · Posted 2005-11-15

Recall StatusTerminated
Product CodeKHO
Date Initiated2005-09-29
Date Terminated2007-07-31
Product Quantity3,507 Vitros ECi units distributed

Reason for Recall

When using Vitro ECi Immunodiagnostic System with Vitros Anti-HBc IgM Reagent Packs and Anti-HAV IgM Reagent Packs, results of the 1st of two required re-test replicates may be negtively biased.

Firm Action

Letters dated 9/29/2005 issued to all domestic customers of the VITROS ECi/ECiQ systems and to Ortho/J&J foreign affiliates. Customers instructed to immediately discontinue using reflex processing to the VITROS Anti-HBc IgM and VITROS Anti-HAV reagent packs.

Root Cause

Other

Distribution Pattern

Nationwide and worldwide distribution of the VITROS ECi Immunodiagnostic systems.