Terminated

John Bunn 1392 Series Jackson Tracheostomy Stainless Steel Tube.

FDA Recall Z-0139-2010 · Posted 2009-11-09

Recall StatusTerminated
Product CodeBTO
Date Initiated2009-09-14
Date Terminated2010-12-30
Product Quantity396 units

Reason for Recall

Incorrect material. Manufacturer changed the material of the trach tube from Stainless Steel to German silver/nickel, contrary to product specifications.

Firm Action

GF Health Products (Graham Field) notified consignees by certified US Mail on/about 09/14/2009. The distribution to the dealers includes the announcement of the recall to the dealer the letter to send to the end users and a Proof of Compliance card for the dealer to submit to Graham Field following completion of the recall instructions. Graham-Field instructed the dealers to return any impacted products on site, as well as any delivered by end-users, customers or other impacted third parties. Dealer will be compensated for executing the recall. Graham Field does not have direct access to end user names; however it plans to reach end-users by financially supporting a mailing from dealers to their individual customers notifying them of this recall. A follow up 60 day certified mailing will be performed for non-responders. The dealers will receive financial support for this second mailing. A secure third company will be contracted to hold for destruction all products affected by this recall. The method of destruction will be to cut the products into pieces, make them not usable. The FDA will be notified prior to the destruction of product and detailed logs and records will be maintained. Periodic effectiveness status report will be maintained for this recall.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide distribution.

Official Source

View FDA Recall →