Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System.
FDA Recall Z-0139-2011 · Posted 2010-10-25
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeCEM
Date Initiated2010-08-19
Date Terminated2010-10-26
Product Quantity3,706 cartons
Reason for Recall
High potassium content in the affected lots results in the potential for low potassium recovery for Dilution Check, QC and patient samples when run on the Dimension Vista System.
Firm Action
Siemens sent an Urgent Field Safety Notice (dated "August 2010") to customers identifying the product, the problem and the action to be taken. Customers were instructed on how to continue testing until new lots of the recalled product were available and asked to complete and fax a Field Correction Effectiveness Check to 301-631-8467 to confirm receipt of the Urgent Field Safety Notice. For technical questions or concerns customers were instructed to contact the Siemens Technical Solutions Center at 800-441-9250.
Root Cause
Other
Distribution Pattern
USA, Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement