Philips BV Pulsera R2.2 Mobile X-ray system. The Type Product Code 718094 can be found under the X-ray trolley cover on the front lower right side of the module or under the stand cover. The device's brochure is labeled in part: "Philips...Mobile C-Arm systems BV Pulsera".
FDA Recall Z-0140-2011 · Posted 2010-10-25
Advertisement
Recalling FirmPhilips Medical Systems North America Co. Phillips
Recall StatusTerminated
Product CodeJAA
Date Initiated2008-09-05
Date Terminated2010-12-07
Product Quantity1331 units
Reason for Recall
Reliability of the Stand-Trolley cable may result in loss of images on the monitor during an examination. This intermittent problem can lead to System lock-up, where no radiation is possible and images are not visible on the monitor
Firm Action
On 10/04/2008, Philips began sending the URGENT-Device Correction BV Pulsera R2.2 and Endura R2.2 letters to their customers.
On 09/03/2008, the firm issued the Field Change Order (FCO) 71800024 and the firm's representative will replace the stand-trolley cable free of charge.
Customers should contact Philips Call Center at 1-800-722-9377, #5, #2, and reference "FCO 71800024" with any questions.
Root Cause
Component design/selection
Distribution Pattern
Nationwide distribution.