Terminated

S5 Heart-lung machine, Stockert S5 System. Catalogue No: 48-30/40/50-00. The Stockert S5 System is intended to be used during cardiopulmonary bypass for procedures lasting six(6) hours or less.

FDA Recall Z-0142-2014 · Posted 2013-11-04

Recall StatusTerminated
Product CodeDTQ
Date Initiated2013-09-24
Date Terminated2014-07-10
Product Quantity84

Reason for Recall

Sorin Group is recalling certain models of S5 Perfusion Systems due to malfunction of cardioplegia control.

Firm Action

Sorin Group sent an Urgent Field Safety Notice dated September 24, 2013, to all affected customers. The letter informed users of the potential for failing of automatic cardioplegia delivery function and notified them that they would be contacted by the Sorin Service Group to arrange an appointment to update the firmware on site. Customers were asked to complete the Response form to confirm they had received, read and understood the Field Notice. For questions regarding this recall call 303-467-6527.

Root Cause

Software Manufacturing/Software Deployment

Distribution Pattern

Worldwide Distribution - USA including WA, NY, VA, TN, SC, TX, CA, CO, MA, WI, MD, AL, ND, and Puerto Rico and Internationally to Italy, Finland, Germany, Sweden, and Great Britain.