Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A

FDA Recall Z-0142-2022 · Posted None

Recall StatusTerminated
Product CodeKPS
Date Initiated2020-11-11
Date Terminated2023-06-30
Product Quantity3 units

Reason for Recall

A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the diagnostic imaging system failing resulting in rescanning and reinjection of contrast medium.

Firm Action

On November 11, 2020, Canon Medical Systems visited affected consignees and updated affected software. After completion of the software update, the consignee was provided a "Recall field Modification Instruction" document concerning the correction.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution in the states of NV, AR MN.