Power cord for Optional Thermal Printer (For use with Model 3300 COC), Oximetrix 3 SO2/CO Computer, Q-Vue CCO Computer and Q2 CCO/SO2 Computer, List Number 50132-04-05, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.
FDA Recall Z-0143-2010 · Posted 2009-11-09
Advertisement
Recalling FirmHospira Inc
Recall StatusTerminated
Product Codedxg
Date Initiated2009-08-11
Date Terminated2012-04-04
Product Quantity204,999 power cords distributed for all devices
Reason for Recall
Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard.
Firm Action
Hospira initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and a letter to consignees was sent via UPS. The firm will replace affected cords and effectiveness checks will be conducted through tracking of servicing.
Root Cause
Component design/selection
Distribution Pattern
Worldwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.