Storz Equipment, Fluid Management Stand, for Fluid Management Systems, Product Number 26601EQ-KT

FDA Recall Z-0144-2008 · Posted 2007-11-07

Recall StatusTerminated
Product CodeHIG
Date Initiated2007-08-17
Date Terminated2012-02-29
Product Quantity287

Reason for Recall

Outflow Difference: The accuracy of the Equimat displayed value could be influenced by external factors of the cart/stand and may result in a fluctuating or inaccurate reading.

Firm Action

All affected customers were notified via a Customer Letter issued on September 10, 2007 via FedEx overnight mail. The letter describes the problem with the device and the health hazard risk, and provides instructions for users to return the old stand and make arrangements for replacement stands.

Root Cause

Device Design

Distribution Pattern

Nationwide