AXIOM Artis Systems VB23A-VB23H with Perivision functionality, Model Numbers: 5904433,5904649, 5904466, 5904441, 5904656, 7728350, and 7007755 Angiographic X-ray System
FDA Recall Z-0144-2011 · Posted 2010-10-26
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeIZI
Date Initiated2010-09-09
Date Terminated2012-09-29
Product Quantity286 units
Reason for Recall
Sporadic system crash while Perivision is performed.
Firm Action
Siemens issued an Urgent Field Safety Notice to all customers. The letter informed customers of the potential issues and provided additional information on the implemented resolution. To resolve the problem, a system update was to be installed on each affected system under Update Instructions AX054/10/S.
For questions phone (610) 448-3237.
Root Cause
Software design
Distribution Pattern
USA - Nationwide Distribution