Fresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.

FDA Recall Z-0148-2014 · Posted 2013-11-05

Recall StatusTerminated
Product CodeFKX
Date Initiated2013-10-14
Date Terminated2021-05-18
Product Quantity16,600

Reason for Recall

Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitis

Firm Action

Fresenius Medical Care contacted all affected consignees by telephone on 10/14/13 to inform them of the product safety alert. Follow-up Urgent Safety Alert letter was sent on 10/18/13 and 10/23/13 by Certified Mail, Return Receipt requested. The letter identified the affected product, problem and actions to be taken. Customers were instructed to report any events associated with this safety alert to Customer Service Team at 1-800-323-5188.

Root Cause

Device Design

Distribution Pattern

USA Nationwide Distribution