Terminated

Misys Laboratory System, Versions 5.2, 5.23, 5.3 up to 5.3.2.

FDA Recall Z-0151-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeJQP
Date Initiated2003-07-07
Date Terminated2004-09-23
Product Quantity473

Reason for Recall

Software defect causes orders to be lost under certain conditions of use. A typographical error at the Accept/Modfy/Reject prompt will cause the order not to be processed by the system.

Firm Action

Product Safety Notice PSN-03-L32 was issued by fax to customers on July 7, 2003. It communcated a procedural workaround and that a follow-up notification was pending completion of a software engineering project to correct the defect.

Root Cause

Other

Distribution Pattern

Nationwide and to United Kingdom, Ireland, Denmark, Canada, Bermuda, Saudi Arabia, United Arab Emirates,