Terminated

HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients

FDA Recall Z-0151-2016 · Posted 2015-11-04

Recall StatusTerminated
Product CodeCBK
Date Initiated2014-04-22
Date Terminated2017-08-24
Product Quantity1128

Reason for Recall

Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult).

Firm Action

Hamilton Medical issued a letter dated April 22, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that their Hamilton Medical Account Manager would contact them to coordinate the deactivation of the Ventilation Suppression feature. Customers were asked to review the Medical Device Field Safety Corrective Action. Customers with questions were instructed to call 800-426-6331, Ext 215.

Root Cause

Device Design

Distribution Pattern

Nationwide Distribution