Urgent Medical Device Voluntary Recall notification letters dated 9/20/19 were sent to customers. An Urgent Medical Device Voluntary Recall REVISED RECALL NOTICE notification letters were sent to customers on 9/27/19.
1. Immediately check your inventory and quarantine affected stock on hand.
2. The following process shall be undertaken in order to segregate any affected
inventory. The existing inventory shall be inspected and any affected product
quarantined and destroyed.
A. Inspect the Tyvek pouch seal integrity.
B. Secondly, the product contained within the pouch may be partially sealed
into the Tyvek seal itself.
3. unaffected product shall be distributed with no futher action necessary.
4. For any product that is quarantined and must be destroyed:
A. Xodus Medical requires documentation of the affected inventory that is
found to be affected and destroyed. Information regarding the following
shall be documented and provided to Xodus Medical.
B. Product Number; Quantity Affected; Method of Destruction; Evidence of Destruction. It can be photos or videos of product being
destroyed
5. Please immediately complete the attached Acknowledgement form even if you do not have any affected stock, and return the acknowledgement to Xodus
Medical at the following: Email
[email protected] or Fax +1 724-337-1131
For further information and question, please call Paul Lloyd, Vice President, Global QA/RA & Technology, Xodus Medical; +1 724-337-5500;
[email protected]
Urgent: Medical Device Voluntary Recall REVISED RECALL NOTICE dated 6/8/20 were sent to customers.
Xodus Medical has received complaints of some devices where the sterile barrier of some devices may have been affected. Xodus Medical has confirmed instances where the pouch seal was compromised on some of the reported product. In no instance has it been reported to Xodus Medical that a compromise in the sterile barrier has resul