Recall StatusOpen, Classified
Date Initiated2023-10-13
Product Quantity2023 units
Reason for Recall
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
Firm Action
Baxter issued an "Urgent Medical Device Correction" notice on 10/13/2023 via USPS first-class mail. The notice explained the issue, hazard involved, and provided mitigation procedures in accordance with the Operator's Manual until a software upgrade becomes available.
For additional questions or experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867, Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide Distribution to states of: AR, AZ, CA, FL, GA, HI, IA, IN, KS, LA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, PA, TX, UT, VA, VT, WA, WI AND WV.