ISE REF Valve, Part No. MU3226 for the AU400, AU640, AU2700, and AU5400 Chemistry Analyzer. Automated chemistry analyzers that measures analytes in samples.
FDA Recall Z-0152-2014 · Posted 2013-11-05
Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeJJE
Date Initiated2013-09-09
Date Terminated2014-09-30
Product Quantity4584 units total (1173 units in US)
Reason for Recall
The recall was initiated because Beckman Coulter has confirmed that some AU Clinical Chemistry Analyzer ISE units can experience premature ISE REF (Reference) valve failure. Premature valve failure can be detected by the presence of air bubbles in the ISE REF solution tubing on the ISE unit. If air bubbles are present, accuracy of the ISE test results may be affected.
Firm Action
A customer notification letter dated 9/9/13 was sent to customers who received the ISE REF Valve, Part Number MU3226 for the AU400, AU640, AU2700, AU5400 and the ISE REF Valve, Part Number MU7638 for the AU480, AU680, AU5800 Chemistry Analyzers. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Beckman Coulter Customer Support Center via http://www.beckmancoulter.com/customersupport/support, via phone at (800) 854-3633 in the US and Canada. Outside of the US and Canada, customers are instructed to contact their local Beckman Coulter Representative.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution-USA (nationwide) and the countries of Canada, Vietnam, Zambia, and Zimbabwe.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K011720OLYMPUS AU5400 CLINICAL CHEMISTRY ANALYZEROlympus America, Inc. · 2001-06-25K003721MODIFICATION TO OLYMPUS AU2700 CLINICAL CHEMISTRY ANALYZEROlympus America, Inc. · 2000-12-18K981743OLYMPUS AU400 CLINICAL CHEMISTRY ANALYZEROlympus America, Inc. · 1998-06-09K961274OLYMPUS AU600 CLINICAL CHEMISTRY ANALYZEROlympus America, Inc. · 1996-07-05