Precision Medical- Easy Mate Portable Liquid Oxygen System (container), PM22010 Series, Model number 2201, Precision Medical, Inc., Northampton, PA 18067
FDA Recall Z-0153-2007 · Posted 2006-11-04
Recalling FirmPrecision Medical, Inc.
Recall StatusTerminated
Product CodeBYJ
Date Initiated2006-10-04
Date Terminated2008-02-15
Product Quantity550 units
Reason for Recall
Component assembly; The fill chuck is not assembled properly which may allow the device to come apart over time releasing its contents.
Firm Action
The recalling firm telephoned the company to inform them of the problem and to make them aware that replacements would be made available starting on 10/16/06.
Root Cause
Other
Distribution Pattern
The product was shipped to home care providers nationwide