Terminated

Precept Medical Products, Tape Fog Shield Surgical Mask Green Fiberglass Free, REORDER #: 65-3322, non-sterile, 50/box, 5 box/case, Precept Medical Products, Inc., Arden, NC 28704, 1-800-438-5827, Mexico

FDA Recall Z-0153-2008 · Posted 2007-12-28

Recall StatusTerminated
Product CodeFYA
Date Initiated2007-09-14
Date Terminated2009-09-11
Product Quantity11,546,350 (both products-Z-0153/0154-2008)

Reason for Recall

Small slits under the folds of the masks; thus they may not provide the protection necessary between the patient and surgical staff

Firm Action

The firm notified its consignees on 09/14/2007 by Urgent Voluntary Medical Device Recall letter (USPS mail, certified, return-receipt requested). The recall was expanded on 10/04/2007 by Urgent Voluntary Medical Device Recall--Expansion Letter to include additional lots of product. Recall letters instruct users to cease use of devices and return them as well as conduct sub-recalls to the users level for any devices they have sub-distributed.

Root Cause

Process control

Distribution Pattern

Worldwide; USA, Australia and Canada