Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Preat REF 9007123, Neodent GM x 4mm Healing Abutment, Rx Only, Non-Sterile, Single Use Only. Dental

FDA Recall Z-0153-2024 · Posted 2023-10-24

Recalling FirmPreat Corp
Recall StatusOpen, Classified
Product CodeNHA
Date Initiated2023-07-18
Product Quantity23 units

Reason for Recall

The healing abutments contain a dimensional condition that may lead to a gap between the healing abutment and the implant allowing for microleakage

Firm Action

On August 4, 2023, Preat Corporation issued a "Urgent Medical Device Recall" notification via: USPS. Preat ask consignees to take the following actions: Product within the scope of the recall shall be discarded or returned to Preat Corporation. Instructions for the return will be provided. Complete and email response form to Nichole Fetterman, E-Mail: [email protected] or mail to Preat Corporation, Attn. Nichole Fetterman, 2625 Skyway Drive, Ste. B, Santa Maria, CA 93455

Root Cause

Process change control

Distribution Pattern

US: FL, CA, OR, TX, OH, CA, SC, OR, NJ, SC, OUS: None

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑