Terminated

Tapered Screw-Vent Implant, MTX, 4.1mmD X 10mmL X 3.5mm Platform Part No. TSV4B10 Tapered Screw-Vent Implant, MTX, 4.1mmD X 11.5mmL X 3.5mm Platform Part No. TSV4B11 The Tapered Screw-Vent Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth

FDA Recall Z-0154-2015 · Posted 2014-10-30

Recalling FirmZimmer Dental Inc
Recall StatusTerminated
Product CodeDZE
Date Initiated2014-10-01
Date Terminated2015-04-28
Product Quantity1,044 units

Reason for Recall

Zimmer Dental is recalling Tapered Screw-Vent Implants because the incorrect diameter size was printed on the inner vial.

Firm Action

Customer Notification letters dated 10/6/14 were sent to all customers who purchased the Tapered Screw Vent Implant and the Spline Twist Implant because the incorrect diameter size was printed on the inner vial. The letters informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the attached Business Reply Form and fax it to (574) 372-4265 or email to [email protected]. Customers with questions are instructed to contact Zimmer Dental Customer Service at (800) 854-7019 from 7am-5pm, Monday-Friday.

Root Cause

Labeling Change Control

Distribution Pattern

Worldwide Distribution-US (nationwide) and the countries of Poland, Canada, Belgium, Italy, France, Israel, Morocco, Spain, Costa Rica, and Japan.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →