ORBIT PT. The ORBIT PT works in conjunction with an TRUMPF mobile operating table to automate the patient transport process.
FDA Recall Z-0155-2014 · Posted 2013-11-05
Advertisement
Recalling FirmTrumpf Medical Systems
Recall StatusTerminated
Product CodeJEA
Date Initiated2013-07-24
Date Terminated2014-06-03
Product QuantityOne
Reason for Recall
An odor and smoke developed - The cause was determined to be a power supply unit sitting in the controller column.
Firm Action
The user facility was contacted via phone in early August. They were contacted 9/17/2013 to schedule the replacement of the power pack because the new power pack became available.
Root Cause
Component design/selection
Distribution Pattern
US Distribution: NJ only.