Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085

FDA Recall Z-0156-2024 · Posted 2023-10-24

Recall StatusOpen, Classified
Product CodeIWE
Date Initiated2023-02-28
Product Quantity52 units

Reason for Recall

An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.

Firm Action

Omega Medical Imaging issued a Field Safety Advisory Notice to its consignees on 02/28/2023 by email. The notice explained the issue and pledged to evaluate all potentially impacted devices and mitigate the issue to bring the boom systems up to the latest safety design and standard of the newer models approved for field use.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Pakistan.