The affected device is labeled Angio-Seal Vascular Closure Device 6F VIP Platform, model 610130. The recalled product is the 6 French model size.

FDA Recall Z-0157-2007 · Posted 2006-11-07

Recall StatusTerminated
Product CodeMGB
Date Initiated2006-06-07
Date Terminated2006-11-13
Product Quantity4,770 Units

Reason for Recall

The Angio Seal VIP 6F devices were incorrectly packaged with a 0.038'' guidewire vs. the required 0.035'' guidewire.

Firm Action

All consignees were notified by letter on June 7, 2006. All product is to be retreived by recalling firm''s sales force for return to St. Jude Medical, in Minnetonka, MN.

Root Cause

Other

Distribution Pattern

Nationwide.