Terminated

ICCG AC-Powered Adjustable Medical Bed

FDA Recall Z-0158-06 · Posted 2005-11-17

Recalling FirmInvacare Corporation
Recall StatusTerminated
Product CodeFNL
Date Initiated2004-07-23
Date Terminated2012-04-12
Product Quantity14,343 units

Reason for Recall

Component Defect. Metal Oxide Varister (MOV) component of the junction box may short circuit with the potential for device failure.

Firm Action

The manufacturer notified their distributors of the recall with telephone calls & a letter. The distributors were provided information about the recall and informed of their recall responsibilites.

Root Cause

Other

Distribution Pattern

World wide