Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Medline brand medical procedure convenience kits, labeled as: a) CONE BIOPSY PACK, REF DYNJ17534L; b) BIOPSY TRAY, REF DYNJ43076A; c) LOU LOOP RECORDER PACK, REF DYNJ49404C; d) CT BIOPSY, REF DYNJ52666C; e) BIOPSY KIT, REF DYNJ61130B; f) ASPEN SOFT TISSUE PACK, REF DYNJ63165B; g) BIOPSY TRAY FOAKS, REF DYNJ63575A; h) CUSTOM SOFT TISSUE PACK, REF DYNJ66347A; i) KWIK PACK, REF DYNJ67926B; j) ESI TRAY, REF DYNJ68261B; k) LARGE BIOPSY PACK NTX, REF DYNJ68393A; l) LOOP KIT, REF DYNJ84129

FDA Recall Z-0158-2025 · Posted 2024-10-23

Recall StatusOpen, Classified
Product CodeOFG
Date Initiated2024-09-04
Product Quantity2,426 units

Reason for Recall

It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.

Firm Action

Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 09/05/2024 via email and USPS. The notice explained the issue, hazard, and requested the destruction of the affected product. Distributors were directed to notify their consignees.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑