Jostra HL-20 Heart Lung Machine - guiding pins
FDA Recall Z-0159-2007 · Posted 2006-11-07
Advertisement
Recalling FirmMAQUET Inc.
Recall StatusTerminated
Product CodeDTQ
Date Initiated2006-10-03
Date Terminated2007-12-19
Product Quantity227 HL-20 base units with 545 single pumps
Reason for Recall
The HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in the tube guide rolls, they cannot roll smoothly and finally they can get stuck.
Firm Action
An 'Important Safety Notice - Please Read Immediately' letter was sent to all US customers on 10/3/06 by FedEx Return Receipt.
Root Cause
Other
Distribution Pattern
Worldwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.