Terumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a 5-digit code beginning with the digit 6: Catalog Number: 6XXXX ; Catalog Codes: Kits with a 5-digit code beginning with the digit 7: Catalog Number: 7XXXX Product usage: The Cardioplegia Delivery Sets are indicated for delivery of cardioplegic solutions to the heart during cardiopulmonary bypass surgery. The Cardiovascular Procedure Kits are indicated for use only in the extracorporeal circuit for the cardiopulmonary bypass procedure(s) for which the user designed it.
FDA Recall Z-0159-2015 · Posted 2014-10-31
Recalling FirmTerumo Cardiovascular Systems Corp
Recall StatusTerminated
Product CodeOEZ
Date Initiated2011-03-01
Date Terminated2014-11-06
Product QuantityCatalog Number: 6XXXX: 4561 units; Catalog Number: 7xxxx : 19356 units
Reason for Recall
Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line.
Firm Action
Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions:
Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.