ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.
FDA Recall Z-0160-2018 · Posted None
Advertisement
Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeHAW
Date Initiated2014-11-06
Date Terminated2020-07-13
Product Quantity31 units
Reason for Recall
Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).
Firm Action
All 9 domestic units were corrected by the firm in February 2017, via technician visits to consignee sites.
Root Cause
Device Design
Distribution Pattern
US distribution in AR, GA, OH, MI, and TX.