Mammomat Revelation with software version VC10 and Biopsy Option Intended Use: For mammography exams, screening, diagnostics, biopsies and dual energy procedures
FDA Recall Z-0160-2020 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeMUE
Date Initiated2019-07-23
Date Terminated2020-05-06
Product Quantity139 WW (33 US)
Reason for Recall
An error can occur when performing a biopsy using the InSpect function. The acquisition workstation may become unresponsive to normal user interaction, can only take place in rare cases
Firm Action
Siemens issued the customer notification on 7/23/19 via XP018/19/S. The Advisory Notice provides the problem, health risk and action to take: provided steps to the operator on how to easily release patient and proceed with exam. Siemens is developing a corrective action to resolve the issue, which will be released via Update Instruction XP021/19/S.
Root Cause
Software design
Distribution Pattern
Nationwide