Terminated

Varian RV Software Varian Treatment for Non-Varian Linacs Version 6.6.5042 and 6.6.5043; Model Number: H46; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Designed to assist the operator of a radiation therapy device in providing accurate treatment setups for each patient by monitoring set up parameters and preventing the radiation therapy from commencing irradiation while any parameter is out of conformance with the treatment plan.

FDA Recall Z-0164-2009 · Posted 2008-10-22

Recall StatusTerminated
Product CodeIYE
Date Initiated2008-06-02
Date Terminated2011-05-03
Product Quantity18

Reason for Recall

A malfunction within the Varis software might result in a misadministration (underdose). The software does not correctly interpret the number, causing field parameters having decimal values to be incorrect.

Firm Action

Recall initiated on June 2, 2008. A product notification letter was distributed by certified mail to all affected consignees.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution --- including USA and countries of Canada, Austria, France, Germany, South Korea, Sweden, and the UK.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑