Allura X-per FD 10/10 x-ray, and fluoroscopy
FDA Recall Z-0165-2007 · Posted 2006-11-14
Recalling FirmPhilips Medical Systems North America Co. Phillips
Recall StatusTerminated
Product CodeIZI
Date Initiated2006-10-12
Date Terminated2010-09-24
Product Quantity47 devices
Reason for Recall
Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
Firm Action
On 10/12/06 the firm issued a letter to their consignees advising of the problem and how to avoid the problem by using specific image formats. If the system should lock up, consignees are to switch the system off and then restart. The firm will be conducting software upgrades to the devices once software is ready (estimated time for FD 10/10 is November 2006 and estimated time for FD 20/10 is January 2007). Later, the firm identified 8 additional units and issued a letter on 12/11/06 to affected customers.
Root Cause
Other
Distribution Pattern
Nationwide.